New page: done: 1. New 4-beam fiber explorer gui with photometric channels(PC) On detector control tab, one can change PC position on detector. Normally one don't need to change the numbers unles... |
No edit summary |
||
Line 1: | Line 1: | ||
done: | done: | ||
1. New 4-beam fiber explorer gui with photometric channels(PC) | 1. New 4-beam fiber explorer gui with photometric channels(PC) | ||
On detector control tab, one can change PC position on detector. Normally one don't need to change the numbers unless PC are realigned. | On detector control tab, one can change PC position on detector. | ||
On fiber mapping tab, one can use the new function (perform map (xchan)) to do fiber exploring for all four together. In case the S/N is too low, one may switch back to old function (perform map), which does fiber exploring one at a time. | Normally one don't need to change the numbers unless PC are realigned. | ||
On fiber mapping tab, one can use the new function (perform map (xchan)) | |||
to do fiber exploring for all four together. In case the S/N is too low, | |||
one may switch back to old function (perform map), which does fiber | |||
exploring one at a time. | |||
2. fixed home function for fiber stages. | 2. fixed home function for fiber stages. | ||
Now if there is a power outage, what one has to do is: first clicked 'home all and index'; second click 'recall last positions'. You should be able to see some lights on the GUI, then do a detailed fiber explorer and may need to change z position a little if the focus is not good. | Now if there is a power outage, what one has to do is: first | ||
If after the two clicks, you still don't see lights, then maybe a new fiber stage's home trigger is broken. Right now these stages have broken home trigger: 1x, 1z, 2x, 4x, 4y(from the gui). You can check this by click 'home all' button, if the position number doesn't go to 0, the corresponding fiber stage has a broken home trigger. If you find a new x or y stage is broken, then just do a wide fiber explorer to find light. If it is z stage (you are in trouble!), you need to search z position step by step. My suggestion is to make the z step 100micron, and search xy range 400*400micro with 20 micron step from the middle (x=64000, y=64000) | clicked 'home all and index'; second click 'recall last positions'. | ||
You should be able to see some lights on the GUI, then do a detailed | |||
fiber explorer and may need to change z position a little if the focus | |||
is not good. | |||
If after the two clicks, you still don't see lights, then maybe a new | |||
fiber stage's home trigger is broken. Right now these stages have broken | |||
home trigger: 1x, 1z, 2x, 4x, 4y(from the gui). You can check this by click | |||
'home all' button, if the position number doesn't go to 0, the corresponding | |||
fiber stage has a broken home trigger. If you find a new x or y stage is | |||
broken, then just do a wide fiber explorer to find light. If it is z stage | |||
(you are in trouble!), you need to search z position step by step. My | |||
suggestion is to make the z step 100micron, and search xy range 400*400micro | |||
with 20 micron step from the middle (x=64000, y=64000) | |||
not yet: | not yet: | ||
1. just start working on the new gui for real time flux,vis and CP. | 1. just start working on the new gui for real time flux,vis and CP. | ||
2. a new button for sending offset to ople | 2. a new button for sending offset to ople |
Revision as of 02:39, 11 June 2010
done: 1. New 4-beam fiber explorer gui with photometric channels(PC)
On detector control tab, one can change PC position on detector. Normally one don't need to change the numbers unless PC are realigned. On fiber mapping tab, one can use the new function (perform map (xchan)) to do fiber exploring for all four together. In case the S/N is too low, one may switch back to old function (perform map), which does fiber exploring one at a time.
2. fixed home function for fiber stages.
Now if there is a power outage, what one has to do is: first clicked 'home all and index'; second click 'recall last positions'. You should be able to see some lights on the GUI, then do a detailed fiber explorer and may need to change z position a little if the focus is not good. If after the two clicks, you still don't see lights, then maybe a new fiber stage's home trigger is broken. Right now these stages have broken home trigger: 1x, 1z, 2x, 4x, 4y(from the gui). You can check this by click 'home all' button, if the position number doesn't go to 0, the corresponding fiber stage has a broken home trigger. If you find a new x or y stage is broken, then just do a wide fiber explorer to find light. If it is z stage (you are in trouble!), you need to search z position step by step. My suggestion is to make the z step 100micron, and search xy range 400*400micro with 20 micron step from the middle (x=64000, y=64000)
not yet: 1. just start working on the new gui for real time flux,vis and CP. 2. a new button for sending offset to ople