Lab alignment procedure
1. make sure the input collimated beam is parallel to table:
1a. put an iris in front of the beam to shrink the laser beam size to ~1mm
1b. iteratively place the same target at two places along the beam, the two places have to be far apart for high precision
and the connection of the two places has to be parallel to the table
1c. adjust the tiptilt of the mirror to make sure the beam hit the same position of the target at the two places. The last
mirror is for fine adjustment for target at farther position, and the second to last mirror is for coarse adjustment for the target
at closer position.
2. make sure the home-made front metal panel is perpendicular to the beam.
2a. assemble the metal pieces, no lens or camera. 2b. open the iris to be about 5 mm 2c. use the edge of a rectangular card to block half of the beam, and let the other half pass through 2d. put a flat mirror tightly against the front metal panel (I used one hand to hold and push the mirror) to reflect the passed-through half beam 2e. loose the four screws that tight the metal baseplate to the linear xyz stage. Rotate the whole metal piece so that the reflected half beam matches the blocked half beam horizontally (meaning they become a circle, but it is not true if they are not vertically aligned) 2f. If the reflected beam and blocked beam don't match vertically, we can put some shims between the triangles panels and the front metal panel. But I didn't find it necessary this time.
3. align the center of the beam to the center of the threaded hole of the front metal panel (rough)
3a. put a second iris close to the camera (theoretically you can first iris, but it is too far away to reach), and beam size small ~1mm. And fully open the first iris. 3b. put the cylindrical target into the hole of the front metal panel. Note, the target doesn't tightly fit into the hole, it can be shaken, that's why this step is rough. 3c. move the second iris to put the laser beam at the center of the target.
4. align the beam with the camera 4a. attach the camera to the front metal panel 4b. switch the light source from laser to white light, and turn on the camera, set exposure to be 0.01s, and keep the second iris small. 4c. the readout size should start with full frame so that you can see where the spot is on the detector. Note: we are in crop mode, so if you see light in a subarray readout, this doesn't mean the light is really in the subarray area. That's why you want to start out with full frame. Then loose the screws on the camera, and move the camera to put the light roughly at the right area (the left up corner is (0,0), ideally we want the center of the light to be at (45,45)). When the light is roughly at the center of the 90X90 array, tight the screws.
5. align the reimaging lens pair with the camera. 5a. there are three images needs to be aligned: the pupil image without any lens in the front; the image with one lense in the front; the pupil image with two lens in the front. If they are not overlapped, I think it is because the optical axis of the lens is not aligned with that of the pupil, as shown in the attached figure.
5b. use the actuators on the linear xyz stage to move the images on the detector. The two pupil images move in the opposite way if you turn the actuators in the same direction. This confirms my understand of the misalignment. So it is possible to align these two images. You just need to iterate a couple of times. 5c. In this case when the two pupil images are align, the image of the focus seems pretty well aligned automatically. If it is not aligned, I haven't thought of how to fix it.