4:10pm. ENG BEAM 1 in line.  WAY OFF on CHAMP TABLE !!
4:10pm. ENG BEAM 1 in line.  WAY OFF on CHAMP TABLE !!
beam1 w1


aligning MIRC and CHAMP to get flux levels.
aligning MIRC and CHAMP to get flux levels.


MIRC BEAM 1.  QUAD 262 XCHAN (pix202): 35  FWHM 7.1 7.1 (using 3mu step)
MIRC BEAM 1.  QUAD 262cts XCHAN (pix202): 35  FWHM 7.1 7.1 (using 3mu step)


4:25pm
4:25pm


ALIGNgui CHAMP:
ALIGNgui CHAMP (250Hz, 50mu):
 
""CHAMP BEAM 1: ""
*great image quality on b1l1.  220cts.
*decent image b1r2. 110cts
 
'''discovered R1/r2 mount was not seating properly.. explaining th eextra wobble we noticed.. ugh. this explains a lot !!'''
 
REALIGNED BEAM 1 using ENG BEAM:
*great image quality on b1l1.  220cts.
*decent image b1r2. 110cts
 
5:08pm
going on to beam2 (s2)
* MIRC beam2: quad 232cts. xchan 28cts (pix215) FWHM 7.1 /8.1 mu (using 3mu step)
 
""pre-alignment""
*CHAMP beam is clearly off the R1 target (before re-alignment)
* ok image quality. not best.  B2 L2 90cts 
* a bit worse image quality B2R3. 70 cts.
* bc image hard to find. image elongated a bit. B2R2 80 cts
 
5:19pm going in lab to align. (done at 5:33p)
 
""after alignment in lab using eng beam"
 
there is some vignetting (maybe 10%) on the PZT. Also, perhaps slight defocus due to mounting of PZT mirror. need to loosen screws and NEED TO REALIGN THIS SUBSYSTEM.
 
*B2R2 100 cts.
*B2R3 70  cts
*B2R2 (it took me forever to find this...but was ok in the end).  85 cts
 
not so much better overall !! strange.
 
6:05pm going to beam 3(s1)
 
* MIRC beam 3: quad 245cts. xchan 21 (pix 221) FWHM. 7.6 / 7.3mu (using 3mu step)


CHAMP BEAM 1
""pre-alignment""
*CHAMP beam.  B3L3: 125 cts image circular

Revision as of 01:45, 16 July 2013


2013July JDM Observing Logs


Goals

  • Install new wolverine2
  • fix mirc spooler bugs that cause loss of ople communication


Punch List

lab --- do super careful alignment of pzt feed on worst beam

check for pzt distortion using engineering beam on all beams

try to improve grounding on mirc?

  • go rid of many peaks by moving mirc ground from the overhead pipe to the champ enclosure which is isolated by 3 chokes. maybe we need more chokes.


turn on vega camera and check interfermetence.

measure flux and xchan ratios for each beam with WL source.

> mirc and champ (?)

check tiptlit centers.

clean up wolverine python.

  • unify mirc branches fb thinks python2.7 code will run fine.
  • cleanup so we have a porpoer svn checkout version running under wolverine /devl and observe is linking to the /devel files.
  • same for champ



Bug/Wish List

  • default values of shuttermatrix
  • get telescope beram order from chara shomehow for mirc and champ, including new buttons on guis.
  • proper slts/prsm labels in fiberexplorer
  • fix defaults (ftgui, mric shutter matrix)
  • mirc: save xy and Z of fiber positions
  • save POPS in mirc header
  • mirkwood server should start w/0 vme on
  • mirkwood spooler should run w/o all servers up (opler/tiptilt, etc)
  • (recent) Delay line positions
  • FIXED fix zeros in uv coordinates FIXED!!
  • send ople a target time instead of 0 0 0 0 0 0 (not needed.. just use the HA which is returned)....
  • mirc and champ: replace pulldown menus and text boxes with other elements.
  • champ: ability to algin in wide field mode.
  • champ. fix stop exposure crap (hard)
  • champ. save syncronized spooler data even for large strokes
  • champ. test phase tracking mode
  • champ filter SPIKES out of amplitude(t) gui plots





Theo Questions

  • client read socket in socket.c .. shouldn't there be sleep in this loop?
  • we use sockman for ople but get same port number to use. normal?

Progress

  • FB found problem with memory alloc in spooler. yay!


on-sky CO-PHASING MIRC+CHAMP

MIRC Ealing Mounts (IR1-6):

  • 67410 77100 57200 84450 54830 87750

CHAMP Ealing Mounts

  • 167630 165589 145564 131554 120755 42920
    • (Wrap-around IR1, ir,2,ir,3,ir4, ir5, bc)

12:08am. HD 177756

  • MIRC OFFSETS
    • (S1 -1.883 S2 -2.264 E1 2.858 E2 1.247 W1 -0.730 W2 0
  • CHAMP BC
    • 6/1 -0.261 1/2 0.026 2/3 0.025 3/4 -0.156 4/5 0.034 5/6 0.332

Observing Plans

YSO MWC 297 CHARA number 320192

  • W1 S2 S1 E1 E2 W2: POPS 2 5 5 1 3 5
  • SRC MWC 297 9:55 -- 11:50
  • CAL 1 HD 167768 9:45--12:00 close cal. new one.
  • CAL 2 HD 177756 10:30--12:30 far away cal used by amber

YSO MWC 361

yes!

2013 Jul 15

4pm. JDM

doing lab alignment of champ using engineering beams

starting with 1. will note the MIRC fluxes and CHAMP fluxes.

4:10pm. ENG BEAM 1 in line. WAY OFF on CHAMP TABLE !!

beam1 w1

aligning MIRC and CHAMP to get flux levels.

MIRC BEAM 1. QUAD 262cts XCHAN (pix202): 35 FWHM 7.1 7.1 (using 3mu step)

4:25pm

ALIGNgui CHAMP (250Hz, 50mu):

""CHAMP BEAM 1: ""

  • great image quality on b1l1. 220cts.
  • decent image b1r2. 110cts

discovered R1/r2 mount was not seating properly.. explaining th eextra wobble we noticed.. ugh. this explains a lot !!

REALIGNED BEAM 1 using ENG BEAM:

  • great image quality on b1l1. 220cts.
  • decent image b1r2. 110cts

5:08pm going on to beam2 (s2)

  • MIRC beam2: quad 232cts. xchan 28cts (pix215) FWHM 7.1 /8.1 mu (using 3mu step)

""pre-alignment""

  • CHAMP beam is clearly off the R1 target (before re-alignment)
  • ok image quality. not best. B2 L2 90cts
  • a bit worse image quality B2R3. 70 cts.
  • bc image hard to find. image elongated a bit. B2R2 80 cts

5:19pm going in lab to align. (done at 5:33p)

""after alignment in lab using eng beam"

there is some vignetting (maybe 10%) on the PZT. Also, perhaps slight defocus due to mounting of PZT mirror. need to loosen screws and NEED TO REALIGN THIS SUBSYSTEM.

  • B2R2 100 cts.
  • B2R3 70 cts
  • B2R2 (it took me forever to find this...but was ok in the end). 85 cts

not so much better overall !! strange.

6:05pm going to beam 3(s1)

  • MIRC beam 3: quad 245cts. xchan 21 (pix 221) FWHM. 7.6 / 7.3mu (using 3mu step)

""pre-alignment""

  • CHAMP beam. B3L3: 125 cts image circular