2011-07-01 (UT) ========== Program: CL6 (YSO Imaging, PI: Millan-Gabet et al.) Observers: Rafael, PJ (S1S2E1E2W1W2: 5-5-1-1-2-5) CLIMB on beam 1,2,3 *** S1W2E1, W2=ref. main goal: try MWC275 with this configuration. LT - 20:30 humidity 30-40% r0 = 5cm set things up on delta sco = HD143275 go for first fringes on HD160915 NIRO ALIGN fringes found at +2800/+5800um 500Hz NONDEST LONGSCAN 21:20 go for first cal HD163955 NIRO ALIGN +2900um/+6100um pixel 31 has ~1/2 the flux of pixel 12 ... NIRO ALIGN again counts 60/40/20 - higher fluxes but still unbalanced ... vis ~ 0.6 r0 = 5-7cm 21:50 trg mwc275 = HD163296 saturating on pixel 12 ... change rate to 1000Hz NONDEST LONGSCAN we see weak fringes *occasionally* on 12 and 23. recorded a few 100 scans. sometimes 23 saturates ... repeat with MEDSCAN to avoid saturation. +2600um/+5900um 12 fringe is very clear, 23 pretty marginal but we think is there +2640um/+5870um saved 100s of scans 22:10 cal HD163955 +2690um/+5570um 22:20 trg mwc275 +2650um/+6000um SNR improving (elevation?), now see fringes much more often, even 23 and even 31! 22:40 cal HD163296 +2720um/+5190um nice data 22:50 trg mwc275 +2620um/+5200um?? in this set we don't see the 23 fringe as much ... 23:10 cal HD169520 +2620um/+4860um lost delay during the acquisition ... 23:30 change to: S2W2E2 (POPs 551) 23:50 check star HD160915 NIRO ALIGN offsets +6940/+3530um r0 = 5-6cm humidity 40% 00:00 cal HD163955 saturating change to 2000Hz NONDEST MEDSCAN +6960um/+3670um 00:25 trg mwc275 +6990um/+3540um nice data, all 3 fringes there most of the time stopped the acquisition after 140 good scans so it doesn't take too long ... 00:55 cal HD163955 +6950um/+3590um 01:00 trg mwc275 +6950um/+3440um ran out of delay with 30 scans to go ... move REF. 01:15 cal HD163955 +6930um/+3500um 01:25 trg mwc275 +6950um/+3380um 01:40 cal HD163955 +6900um/+3450um 01:45 trg mwc275 +6890um/+3330um 01:55 cal HD163955 +6870um/+3400um 02:05 trg mwc275 +6840um/+3250um 02:17 out of delay go for v1295aql = HD190073 02:20 check star HD187691 = CHARA 260525 NIRO ALIGN +7280um/+3060um 1000Hz NONDEST MEDSCAN 02:40 cal HD188385 +7250um/+3050um 02:50 trg v1295aql +7240um/+3020um clearly see 3 fringes, 12 and 23 weak, especially 12 ... stopped after 130 good scans ... try 750Hz see if we get more good fringes ... doesn't really help much ... 03:20 cal HD188385 +7200um/+2879um crap ... took these data at 750Hz by error ... now really back to 1000Hz 03:27 trg v1295aql not really seeing the 12,23 fringes so much anymore ... high SNR 31 fringe ... 03:50 cal HD188385 +7180um/+2756um 03:55 trg v1295aql +7260um/+2860um do we have only the cross-fringe?? 04:05 cal HD189509 +7166um/+2725um 04:30 DONE. END OF RUN. --- Summary: S1W2E1: mwc275 - 3 brackets. S2W2E2: mwc275 - 5 brackets. S2W2E2: V1295 Aql - 3 brackets. ---